Measuring Range and Measuring Error
Range of scanning: 0 mm ~10000 mm
Resolution for scanning: 0.1mm (2.5mm ~100mm)
1mm (100 mm ~5000mm)
Range of gain: 0dB ~120 dB
D-Delay: -20μs~+3400μs
P-Delay: 0μs~99.99μs
Sound speed : 1000 m/s~15000m/s
Operation Environment
Temperature: -10℃~50℃
Humidity: 20%~90%RH
Free of strong magnetic field and corrosion.
Power supply
Li battery 7.4V 4800mAh
Overall Dimension and Weight
Overall dimension: 240mm×150mm×50mm
Weight: 1.6kg
Performance Features
l Automated calibration of transducer Zero Offset and/or Velocity ;
l Automated gain、Peak Hold and Peak Memory;
l Automated display precise flaw location(Depth d、level p、distance s、amplitude、sz dB、ф);
l Automated switch three staff gauge ((Depth d、level p、distance s);
l 100 independence setup, any criterion can be input freely, we can work in the scene without test block;
l Big memory of 300 A graph.
l Gate and DAC alarm;Acoustic-Optical alarm;
l RS232/USB port,communication with pc is easy;
l The embeded software can be online updated;
l Li battery, continue working time up to 10 hours;
l Display freeze;
l Automated echo degree;
l Angles and K-value;
l Lock and unlock function of system parameters;
l Dormancy and screen savers;
l Electronic clock calendar ;
l Two gates setting and alarm indication;
l High-speed capture and very low noise;
l DAC、AVG、B Scan ; Solid metal housing ;
l Automated calculate the size of the flaw with w ide bottom type in AVG function.
l AWS D1.1;
l 6dB DAC functions;
l Automated make video of test process and play;
l Provides high contrast viewing of the waveform from bright, direct sunlight to completae darkness and easy to
read from all angles;
l Powerful pc software and reports can be export to excel;
Technical Parameters
Designation | Technical Data |
Range of scanning (mm) | Range of scanning (mm):0~10000 Steps: 2.5,5,10,20, 30,40,50,60,70,80,90, 100,150,200, 250, 300, 350,400,450,500,600,700,800,900,1000,2000,3000,4000,5000,6000,7000,8000,10000 Adjusting step: 0.1mm(2.5 mm~99.9mm),1mm(100mm~10000mm) |
D-delay (ms) | D-delay (ms):-20~+3400 Steps: -20,-10,0.0, 10, 20, 50,100,150,200,250,300,350,400,450,500, 600, 700,800,900,1000,1500,2000,2500,3000,3400. Adjusting steps: 0.1(-20ms~999.9ms),1(1000ms~3400ms) |
P-delay (ms) | P-delay:0.0~99.99 Adjusting steps: 0.01 |
MTLVEL(m/s) | MTLVEL:1000~15000 7 fixed levels: 2260,2730,3080,3230,4700,5920,6300,12000 Adjusting steps: 1 |
Working mode | Single probe (receiving and sending), double probe (one for receiving and another for sending), transmission (transmission probe) |
Frequency Range (MHz) | 0.5~15 |
Gain adjustment (dB) | 0~120 Adjusting step: 0.0,0.1,0.5,1,2,6,12 |
Reject | 0%~80% of screen height, step: 1% |
Vertical linear error | Vertical linear error is not more than 3% |
Horizontal linear error | Not more than 0.2% in the scanning range |
Sensitivity Leavings | ≥62 dB |
Dynamic range | ≥34dB |
Alarm | Three modes, i.e. forbidden wave, loss wave and auto |
A-Scan display area | Full screen or local A-Scan display freezing and de-freezing A-Scan filling |
Data save | 300 A-Scan images (including setting of instrument) |
Standard communication interface with PC | RS232/USB |
Measuring unit | mm/inch |
Battery | Li battery 7.4V 4800mAh |
Power adaptor | Input 100V~240V/50Hz~60Hz Output 9VDC/1.5A |
Working temperature | -10℃~50℃ |
Working humidity | 20%~90% |
Port type | BNC / LEMO |
Overall dimension (mm) | 240×150×50 |
Weight (kg) | 1.6 |
Standard Configurations and Options
Standard Configuration
Table 1.1 List of standard configurations
Designation | Quantity |
Main unit | 1 |
1.5A/9V power adaptor | 1 |
probe connecting cable | 2 |
Carrying case | 1 |
Instruction manual | 1 |
Straight probe | 10mm 4MHz (One) |
Angle probe | 8×9 60°4MHz (One) |
Options
Table 1.2 List of options
Designation | Quantity |
Serial cable | 1 (9 pin) |
Communication software for PC | 1 set |
Send an inquiry for this product