HS600 Ultrasonic Flaw Detector
HS600 Ultrasonic Flaw Detector

HS600 Ultrasonic Flaw Detector

  • This is a portable industrial non-destructive flaw detector, which can rapidly, easily and accurately inspect,locate, evaluate and diagnose various defects (crack, inclusion and pinhole, etc.) in a workpiece without destruction.It can be used both in a lab and field. The instrument can be widely used in any fields that need defect inspection and quality controlling e.g. manufacturing industry, iro

Measuring Range and Measuring Error

Range of scanning: 0 mm ~10000 mm

Resolution for scanning: 0.1mm (2.5mm ~100mm)

1mm (100 mm ~5000mm)

Range of gain: 0dB ~120 dB

D-Delay: -20μs+3400μs

P-Delay: 0μs99.99μs

Sound speed : 1000 m/s15000m/s

Operation Environment

Temperature: -10~50

Humidity: 20%90%RH

Free of strong magnetic field and corrosion.

Power supply

Li battery 7.4V 4800mAh

Overall Dimension and Weight

Overall dimension: 240mm×150mm×50mm

Weight: 1.6kg

Performance Features

l Automated calibration of transducer Zero Offset and/or Velocity

l Automated gainPeak Hold and Peak Memory

l Automated display precise flaw location(Depth dlevel pdistance samplitudesz dBф)

l Automated switch three staff gauge ((Depth dlevel pdistance s)

l 100 independence setup, any criterion can be input freely, we can work in the scene without Test block

l Big memory of 300 A graph.

l Gate and DAC alarmAcoustic-Optical alarm;

l RS232/USB portcommunication with pc is easy

l The embeded software can be online updated

l Li battery, continue working time up to 10 hours;

l Display freeze;

l Automated echo degree;

l Angles and K-value;

l Lock and unlock function of system parameters;

l Dormancy and screen savers;

l Electronic clock calendar ;

l Two gates setting and alarm indication;

l High-speed capture and very low noise;

l DACAVGB Scan ; Solid metal housing ;

l Automated calculate the size of the flaw with w ide bottom type in AVG function.

l AWS D1.1

l 6dB DAC functions;

l Automated make video of test process and play;

l Provides high contrast viewing of the waveform from bright, direct sunlight to completae darkness and easy to

read from all angles;

l Powerful pc software and reports can be export to excel;

Technical Parameters

Designation

Technical Data

Range of scanning (mm)

Range of scanning (mm)010000

Steps: 2.5,5,10,20, 30,40,50,60,70,80,90, 100,150,200, 250, 300, 350,400,450,500,600,700,800,900,1000,2000,3000,4000,5000,6000,7000,8000,10000

Adjusting step: 0.1mm2.5 mm99.9mm),1mm100mm10000mm

D-delay (ms)

D-delay (ms)-20+3400

Steps: -20,-10,0.0, 10, 20, 50,100,150,200,250,300,350,400,450,500, 600,

700,800,900,1000,1500,2000,2500,3000,3400.

Adjusting steps: 0.1-20ms999.9ms),11000ms3400ms

P-delay (ms)

P-delay0.099.99

Adjusting steps: 0.01

MTLVELm/s

MTLVEL1000~15000

7 fixed levels: 2260,2730,3080,3230,4700,5920,6300,12000

Adjusting steps: 1

Working mode

Single probe (receiving and sending), double probe (one for receiving and another for sending), transmission (transmission probe)

Frequency Range (MHz)

0.5~15

Gain adjustment (dB)

0120

Adjusting step: 0.00.10.512612

Reject

0%80% of screen height, step: 1%

Vertical linear error

Vertical linear error is not more than 3%

Horizontal linear error

Not more than 0.2% in the scanning range

Sensitivity Leavings

≥62 dB

Dynamic range

≥34dB

Alarm

Three modes, i.e. forbidden wave, loss wave and auto

A-Scan display area

Full screen or local

A-Scan display freezing and de-freezing A-Scan filling

Data save

300 A-Scan images (including setting of instrument)

Standard communication interface with PC

RS232/USB

Measuring unit

mm/inch

Battery

Li battery 7.4V 4800mAh

Power adaptor

Input 100V240V/50Hz60Hz

Output 9VDC/1.5A

Working temperature

-10℃~50℃

Working humidity

20%~90%

Port type

BNC / LEMO

Overall dimension (mm)

240×150×50

Weight (kg)

1.6

Standard Configurations and Options

Standard Configuration

Table 1.1 List of standard configurations

Designation

Quantity

Main unit

1

1.5A/9V power adaptor

1

probe connecting cable

2

Carrying case

1

Instruction manual

1

Straight probe

10mm 4MHz (One)

Angle probe

8×9 60°4MHz (One)

Options

Table 1.2 List of options

Designation

Quantity

Serial cable

1 (9 pin)

Communication software for PC

1 set

Send an inquiry for this product